Remote Control Commands
R&S
®
SMA100B
500User Manual 1178.3834.02 ─ 05
Example: Performing a power measurement with a fixed filter
Prerequisite: The sensor is connected to the instrument and mapped to the first sensor
channel.
SENSe1:SOURce RF
//Sensor measures the power of the RF signal
SENSe1:FILTer:TYPE NSRatio
//Selects fixed noise filter mode
SENSe1:FILTer:NSRatio 0.02 DB
//Sets the maximum noise component in the result to 0.02 DB
SENSe1:FILTer:NSRatio:MTIMe 10
//Limits the settling time to 10 seconds.
:SENSe1:APERture:DEFault:STATe 0
// Deactivates the default aperture time of the sensor
:SENSe1:APERture:TIMe 10e-6
// Sets the aperture time to 10 us
SENSe1:UNIT DBM
//Selects unit dBm for the measured value
:INITiate:CONTinuous ON
//Switches the continous power measurement on
:READ?
//Triggers the measurement and displays the results
:SLISt[:LIST]?................................................................................................................ 501
:SLISt:SCAN[:STATe]......................................................................................................501
:SLISt:ELEMent<ch>:MAPPing........................................................................................501
:SLISt:SENSor:MAP.......................................................................................................502
:INITiate<hw>[:POWer]:CONTinuous............................................................................... 502
:READ<ch>[:POWer]?....................................................................................................502
:SENSe<ch>:UNIT[:POWer]............................................................................................503
:SENSe<ch>[:POWer]:APERture:DEFault:STATe.............................................................. 503
:SENSe<ch>[:POWer]:APERture:TIMe.............................................................................504
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect.......................................................504
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe........................................................504
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?.........................................................505
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority.........................................................505
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe........................................................... 505
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?.................................................................. 505
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER].................................................................. 506
:SENSe<ch>[:POWer]:FILTer:NSRatio............................................................................. 506
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe.................................................................. 507
:SENSe<ch>[:POWer]:FILTer:SONCe.............................................................................. 507
SENSe, READ, INITiate and SLISt Subsystems