EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Page 1002

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1002User Manual 1175.6632.02 ─ 30
Example:
SOURce1:BB:DM:FORMat APSK32
SOURce1:BB:DM:APSK32:GAMMa G9D10
Manual operation: See "Gamma/Gamma 1" on page 274
[:SOURce<hw>]:BB:DM:SWITching:STATe <State>
Enables switching between a modulated and an unmodulated signal.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SCONfiguration:MODE ADV
SCONfiguration:APPLy
SOURce1:BB:DM:SWITching:STATe ON
Manual operation: See "State Modulation CW Switching" on page 274
Power Ramping
[:SOURce<hw>]:BB:DM:PRAMp:SOURce <Source>
Sets the source for the power ramp control signals.
Parameters:
<Source> INTernal
*RST: INTernal
Example: See Example "Enable power ramping" on page 991
Options: R&S SMW-B10
Manual operation: See "Source" on page 277
[:SOURce<hw>]:BB:DM:PRAMp:SHAPe <Shape>
Sets the edge shape of the ramp envelope.
Parameters:
<Shape> LINear | COSine
*RST: COSine
Example: See Example "Enable power ramping" on page 991
Options: R&S SMW-B10
Manual operation: See "Ramp Function" on page 278
[:SOURce<hw>]:BB:DM:PRAMp:TIME <Time>
Sets the power ramping rise time and fall time for a burst.
SOURce Subsystem

Table of Contents