Remote Control Commands
R&S
®
SMW200A
1004User Manual 1175.6632.02 ─ 30
The ON setting is mandatory if, with power ramping active, only the baseband signal is
output (I/Q outputs).
In case of two-path instruments, the ON setting is also mandatory if a baseband signal
is applied to both paths.
Only then can a signal with a defined, predictable level be output.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Enable power ramping" on page 991.
Options: R&S SMW-B10
Manual operation: See "In Baseband Only" on page 278
[:SOURce<hw>]:BB:DM:PRAMp[:STATe]
<State>
Enables or disables power ramping.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Enable power ramping" on page 991
Options: R&S SMW-B10
Manual operation: See "State" on page 277
Trigger Settings
[:SOURce<hw>]:BB:DM[:TRIGger]:SEQuence <Sequence>
Selects the trigger mode. For detailed description of the trigger modes, refer to "Impact
of the Trigger Modes on the Signal Generation" on page 237.
Parameters:
<Sequence> AUTO | RETRigger | AAUTo | ARETrigger | SINGle
*RST: AUTO
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990
Manual operation: See " Mode" on page 253
[:SOURce<hw>]:BB:DM:TRIGger:SOURce <Source>
Selects the trigger signal source and determines the way the triggering is executed.
Provided are:
Internal triggering by a command (INTernal)
External trigger signal via one of the local or global connectors
SOURce Subsystem