Remote Control Commands
R&S
®
SMW200A
1005User Manual 1175.6632.02 ─ 30
– EGT1|EGT2: External global trigger
– EGC1|EGC2: External global clock
– ELTRigger: External local trigger
– ELCLock: External local clock
●
Internal triggering by a signal from the other basebands (INTA|INTB)
●
In primary-secondary instrument mode, the external baseband synchronization sig-
nal (BBSY)
●
OBASeband|BEXTernal|EXTernal: Setting only
Provided only for backward compatibility with other Rohde & Schwarz signal gener-
ators.
The R&S SMW accepts these values and maps them automatically as follows:
EXTernal = EGT1, BEXTernal = EGT2, OBASeband = INTA or INTB
(depending on the current baseband)
Parameters:
<Source> INTB|INTernal|OBASeband|EGT1|EGT2|EGC1|EGC2|ELTRig-
ger|INTA|ELCLock|BEXTernal|EXTernal | BBSY
*RST: INTernal
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990
Options: ELTRigger|ELCLock require R&S SMW-B10
BBSY require R&S SMW-B9
Manual operation: See "Source" on page 254
[:SOURce<hw>]:BB:DM:TRIGger:SLENgth <SLength>
Defines the length of the signal sequence to be output in the SINGle trigger mode.
Parameters:
<SLength> integer
Range: 1 to 4294967295
*RST: 1000
Default unit: symbol
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990
Manual operation: See " Signal Duration" on page 253
[:SOURce<hw>]:BB:DM:TRIGger:RMODe?
Queries the status of signal generation.
Return values:
<RMode> STOP | RUN
SOURce Subsystem