Remote Control Commands
R&S
®
SMW200A
1007User Manual 1175.6632.02 ─ 30
Parameters:
<Delay> float
Range: 0 to 2147483647
Increment: 0.01
*RST: 0
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990
Manual operation: See "(Specified) External Delay/(Specified) Trigger Delay"
on page 255
[:SOURce<hw>]:BB:DM:TRIGger:OBASeband:RDELay?
Queries the time a trigger event form the other path is delayed.
Return values:
<ObResTimeDelSec> float
Range: 0 to 688
Increment: 0.25E-9
*RST: 0
Example: See Example "Specifying delay and inhibit values in time units"
on page 1022
Usage: Query only
Manual operation: See "Actual Trigger Delay/Actual External Delay" on page 256
[:SOURce<hw>]:BB:DM:TRIGger:OBASeband:TDELay <ObasTimeDelay>
Specifies the trigger delay for triggering by the signal from the other path.
Parameters:
<ObasTimeDelay> float
Range: 0 to 7929.170398682
Increment: 250E-12
*RST: 0
Default unit: s
Example: See Example "Specifying delay and inhibit values in time units"
on page 1022.
Manual operation: See "(Specified) External Delay/(Specified) Trigger Delay"
on page 255
[:SOURce<hw>]:BB:DM:TRIGger:OBASeband:INHibit <Inhibit>
Specifies the number of symbols by which a restart is inhibited. This command applies
only for triggering by the second path.
SOURce Subsystem