EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Page 1010

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMW200A
1010User Manual 1175.6632.02 ─ 30
Marker Settings
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:MODE <Mode>
Defines the signal for the selected marker output.
For detailed description of the regular marker modes, refer to "Marker Modes"
on page 232.
Parameters:
<Mode> CLISt | PULSe | PATTern | RATio
CLISt
A marker signal that is defined in the selected control list is gen-
erated.
*RST: RATio
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990
Manual operation: See "Marker Mode" on page 257
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:ONTime <OnTime>
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:OFFTime <OffTime>
Sets the number of symbols in a period (ON time + OFF time) for marker RATio.
*)
If R&S SMW-B9 is installed, the minimum marker duration depends on the sample/
symbol rate.
See "Marker Minimum Duration" on page 234.
Parameters:
<OffTime> integer
Range: 1 (R&S SMW-B10) / 1* (R&S SMW-B9) to
16777215
*RST: 1
Default unit: symbol
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990
Manual operation: See "Marker Mode" on page 257
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:PATTern <Pattern>, <BitCount>
Defines the bit pattern used to generate the marker signal.
Parameters:
<Pattern> numeric
*RST: #H2