Remote Control Commands
R&S
®
SMW200A
1012User Manual 1175.6632.02 ─ 30
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990.
Manual operation: See "Marker x Delay" on page 257
Clock Settings
[:SOURce<hw>]:BB:DM:CLOCk:MODE <Mode>
Sets the type of externally supplied clock.
Parameters:
<Mode> SYMBol
*RST: SYMBol
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990.
Options: R&S SMW-B10
Manual operation: See "Clock Mode" on page 258
[:SOURce<hw>]:BB:DM:CLOCk:SOURce <Source>
Selects the clock source:
●
INTernal: Internal clock reference
●
ELCLock: External local clock
●
EXTernal = ELCLock: Setting only
Provided for backward compatibility with other Rohde & Schwarz signal generators
Parameters:
<Source> INTernal|ELCLock|EXTernal
*RST: INTernal
Example: See Example "Adjusting clock, marker and trigger settings"
on page 990.
Options: ELCLock requires R&S SMW-B10
Manual operation: See "Clock Source" on page 258
Handling List Files
[:SOURce<hw>]:BB:DM:CLISt:CATalog?
[:SOURce<hw>]:BB:DM:FLISt:CATalog?
[:SOURce<hw>]:BB:DM:MLISt:CATalog?
[:SOURce<hw>]:BB:DM:DLISt:CATalog?
Reads out the list files present in the default directory (see :MMEMory:CDIRectory).
SOURce Subsystem