Remote Control Commands
R&S
®
SMW200A
1070User Manual 1175.6632.02 ─ 30
Clock Commands
[:SOURce<hw>]:BB:ARBitrary:CLOCk:SOURce <Source>
Selects the clock source:
INTernal: Internal clock reference
ELCLock: External local clock
EXTernal = ELCLock: Setting only
Provided for backward compatibility with other Rohde & Schwarz signal generators
Parameters:
<Source> INTernal|ELCLock|EXTernal
*RST: INTernal
Example: See Example "Managing waveform files" on page 1020.
Options: ELCLock requires R&S SMW-B10
Manual operation: See "Clock Source" on page 258
[:SOURce<hw>]:BB:ARBitrary:CLOCk:MODE <Mode>
Enters the type of externally supplied clock.
Parameters:
<Mode> SAMPle
*RST: SAMPle
Example: See Example "Managing waveform files" on page 1020
Options: R&S SMW-B10
Manual operation: See "Clock Mode" on page 258
14.19.4.4 SOURce:BB:MCCW Subsystem
This subsystem contains the commands for setting the Multi-Carrier CW signals.
The generation of multi-carrier CW signals requires an instrument equipped with the
software options R&S SMW-K61.
Common Suffixes
The following common suffixes are used in remote commands:
Suffix Value range Description
ENTity<ch>
1 to 4 Entity in a multiple entity configuration
SOURce<ch>
[1] to 4 Available baseband signals
OUTPut<ch>
1 to 3 Available markers
SOURce Subsystem