Remote Control Commands
R&S
®
SMW200A
1083User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:BB:MCCW:TRIGger:EXTernal:SYNChronize:OUTPut <Output>
Enables signal output synchronous to the trigger event.
Parameters:
<Output> 0 | 1 | OFF | ON
*RST: 1
Example: See "Programming Examples" on page 1071
Manual operation: See "Sync. Output to External Trigger/Sync. Output to Trigger"
on page 254
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:DELay <Delay>
Specifies the trigger delay for external triggering.
Parameters:
<Delay> float
Range: 0 to 2147483647
Increment: 0.01
*RST: 0
Default unit: samples
Example: See "Programming Examples" on page 1071
Manual operation: See "(Specified) External Delay/(Specified) Trigger Delay"
on page 255
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:INHibit <Inhibit>
Specifies the number of samples by which a restart is inhibited following a trigger event
from the other path.
Parameters:
<Inhibit> integer
Range: 0 to 67108863
*RST: 0
Example: See "Programming Examples" on page 1071
Manual operation: See "External / Trigger Inhibit" on page 255
[:SOURce<hw>]:BB:MCCW:TRIGger:RMODe?
Queries the status of signal generation for all trigger mode, if multi-carrier CW genera-
tion is on.
Return values:
<RMode> STOP | RUN
Example: See "Programming Examples" on page 1071
Usage: Query only
SOURce Subsystem