Remote Control Commands
R&S
®
SMW200A
1084User Manual 1175.6632.02 ─ 30
Manual operation: See "Running/Stopped" on page 253
[:SOURce<hw>]:BB:MCCW:TRIGger:SLENgth <SLength>
Defines the length of the signal sequence to be output in the "Single" trigger mode.
Parameters:
<SLength> integer
Range: 1 to 1000
*RST: 1
Default unit: samples
Example: See "Programming Examples" on page 1071
Manual operation: See " Signal Duration" on page 253
[:SOURce<hw>]:BB:MCCW:TRIGger:SOURce <Source>
Selects the trigger signal source and determines the way the triggering is executed.
Provided are:
Internal triggering by a command (INTernal)
External trigger signal via one of the local or global connectors
EGT1|EGT2: External global trigger
EGC1|EGC2: External global clock
ELTRigger: External local trigger
ELCLock: External local clock
Internal triggering by a signal from the other basebands (INTA|INTB)
In primary-secondary instrument mode, the external baseband synchronization sig-
nal (BBSY)
OBASeband|BEXTernal|EXTernal: Setting only
Provided only for backward compatibility with other Rohde & Schwarz signal gener-
ators.
The R&S SMW accepts these values and maps them automatically as follows:
EXTernal = EGT1, BEXTernal = EGT2, OBASeband = INTA or INTB
(depending on the current baseband)
Parameters:
<Source> INTB|INTernal|OBASeband|EGT1|EGT2|EGC1|EGC2|ELTRig-
ger|INTA|ELCLock|BEXTernal|EXTernal | BBSY
*RST: INTernal
Example: See "Programming Examples" on page 1071
Options: ELTRigger|ELCLock require R&S SMW-B10
BBSY require R&S SMW-B9
Manual operation: See "Source" on page 254
SOURce Subsystem