Remote Control Commands
R&S
®
SMW200A
1085User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal]:DELay <Delay>
Specifies the trigger delay in samples.
Maximum trigger delay and trigger inhibit values depend on the installed options. See
"Specifying delay and inhibit values" on page 242.
Parameters:
<Delay> float
Range: 0 to 2147483647
Increment: 0.01
*RST: 0
Default unit: samples
E.g. 0 to 2147483647 samples (R&S SMW-B10)
Example: See "Programming Examples" on page 1071.
Manual operation: See "(Specified) External Delay/(Specified) Trigger Delay"
on page 255
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal]:INHibit
<Inhibit>
Specifies the number of samples, by which a restart is inhibited following an external
trigger event.
Maximum trigger delay and trigger inhibit values depend on the installed options. See
"Specifying delay and inhibit values" on page 242.
Parameters:
<Inhibit> integer
Range: 0 to 21.47 * (clock frequency)
*RST: 0
Default unit: sample
E.g. 0 to 2147483647 samples (R&S SMW-B10)
Example: See "Programming Examples" on page 1071
Manual operation: See "External / Trigger Inhibit" on page 255
[:SOURce<hw>]:BB:MCCW[:TRIGger]:SEQuence <Sequence>
Selects the trigger mode. For detailed description of the trigger modes, refer to "Impact
of the Trigger Modes on the Signal Generation" on page 237.
Parameters:
<Sequence> AUTO | RETRigger | AAUTo | ARETrigger | SINGle
*RST: AUTO
Example: See "Programming Examples" on page 1071
Manual operation: See " Mode" on page 253
SOURce Subsystem