Remote Control Commands
R&S
®
SMW200A
1088User Manual 1175.6632.02 ─ 30
Manual operation: See "Marker x Delay" on page 257
Clock Settings
[:SOURce<hw>]:BB:MCCW:CLOCk:SOURce................................................................. 1088
[:SOURce<hw>]:BB:MCCW:CLOCk:MODE.................................................................... 1088
[:SOURce<hw>]:BB:MCCW:CLOCk:SOURce <Source>
Selects the clock source:
●
INTernal: Internal clock reference
●
ELCLock: External local clock
●
EXTernal = ELCLock: Setting only
Provided for backward compatibility with other Rohde & Schwarz signal generators
Parameters:
<Source> INTernal|ELCLock|EXTernal
*RST: INTernal
Example: See "Programming Examples" on page 1071.
Options: ELCLock requires R&S SMW-B10
Manual operation: See "Clock Source" on page 258
[:SOURce<hw>]:BB:MCCW:CLOCk:MODE <Mode>
Sets the type of externally supplied clock.
Parameters:
<Mode> SAMPle
*RST: SAMPle
Example: See "Programming Examples" on page 1071.
Options: R&S SMW-B10
Manual operation: See "Clock Mode" on page 258
14.19.4.5 SOURce:AWGN Subsystem
The SOURce:AWGN subsystem contains the commands for setting the noise generator.
Suffixes in the keywords ENTity<ch> and SOURce<hw>
You can address multiple entities configurations by using the SCPI commands starting
with the keyword SOURce or the alias commands starting with the keyword ENTity.
SOURce Subsystem