Remote Control Commands
R&S
®
SMW200A
1117User Manual 1175.6632.02 ─ 30
Parameters:
<Frequency> Frequency#1[, Frequency#2, ...]
String of values with default unit Hz.
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1114 .
Manual operation: See "Edit List Mode Data" on page 525
[:SOURce<hw>]:CORRection:CSET:DATA:POWer <Power>
Enters the level values to the table selected with [:SOURce<hw>]:CORRection:
CSET[:SELect].
Parameters:
<Power> Power#1[, Power#2, ...]
String of values with default unit dB.
*RST: 0
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1114 .
Manual operation: See "Edit List Mode Data" on page 525
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
Queries the number of frequency/level values in the selected table.
Return values:
<Points> integer
Range: 0 to 10000
*RST: 0
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1114 .
Usage: Query only
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Fills the selected user correction table with the level values measured by the power
sensor for the given frequencies.
To select the used power sensor set the suffix in key word SENSe.
Example:
See Example "Fill user correction data with sensor"
on page 1115.
Usage: Event
Manual operation: See "Fill User Correction Data with Sensor" on page 578
SOURce Subsystem