Remote Control Commands
R&S
®
SMW200A
1119User Manual 1175.6632.02 ─ 30
Example: See Example "Fill user correction data with sensor"
on page 1115.
Usage: Event
[:SOURce<hw>]:CORRection[:STATe] <State>
Activates user correction with the currently selected table.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1114 .
Manual operation: See "State" on page 573
[:SOURce]:CORRection:CSET:CATalog?
Queries a list of available user correction tables.
Return values:
<Catalog> string
List of list filenames, separated by commas
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1114 .
Usage: Query only
Manual operation: See "UCOR Data ..." on page 574
See "Directory, File List and Filename" on page 682
[:SOURce]:CORRection:CSET:DELete <Filename>
Deletes the specified user correction list file.
Setting parameters:
<Filename> string
Filename or complete file path; file extension is optional.
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 1114 .
Usage: Setting only
Manual operation: See "UCOR Data ..." on page 574
14.19.5.2 Correction Data Exchange
With the following commands, you can configure user correction lists and export or
import them accordingly.
SOURce Subsystem