Remote Control Commands
R&S
®
SMW200A
1127User Manual 1175.6632.02 ─ 30
<Mode>
Parameter in manual oper-
ation
Description
COUPled A Internal & A->B Coupled Assigns the internal oscillator signal of path
A also to path B.
ECOupled A External & A->B Coupled Assigns an externally supplied signal to both
paths.
BOFF A Internal & B RF Off Uses the internal local oscillator signal of
path A, if the selected frequency exceeds
the maximum frequency of path B.
EBOFf A External & B RF Off Uses the LO IN signal for path A, if the
selected RF frequency exceeds the maxi-
mum frequency of path B.
AOFF A RF Off & B External Uses the LO IN signal for path B, if the
selected RF frequency exceeds the maxi-
mum frequency of path A.
Parameters:
<Mode> INTernal | EXTernal | COUPled | ECOupled | BOFF | AOFF
See Table 14-11
*RST: INTernal
Example: See Example "Set up of LO coupling and LO level adjustment
with three Rohde & Schwarz signal generators" on page 1124.
Manual operation: See "Mode" on page 485
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:FREQuency?
Queries the current frequency of the local oscillator at the [LO OUT] connector.
Return values:
<Frequency> float
Range: 100E3 to 20E9
Increment: 0.01
*RST: 0
Example: See Example "Set up of LO coupling and LO level adjustment
with three Rohde & Schwarz signal generators" on page 1124.
Usage: Query only
Manual operation: See "LO OUT Frequency" on page 486
[:SOURce<hw>]:FREQuency:LOSCillator:OUTPut:STATe <State>
Activates the LO output in the second path.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
SOURce Subsystem