Remote Control Commands
R&S
®
SMW200A
1128User Manual 1175.6632.02 ─ 30
Example: See Example "Set up of LO coupling and LO level adjustment
with three Rohde & Schwarz signal generators" on page 1124.
Manual operation: See "Out State " on page 486
[:SOURce<hw>]:FREQuency:MODE <Mode>
Sets the frequency mode for generating the RF output signal. The selected mode
determines the parameters to be used for further frequency settings.
Parameters:
<Mode> CW | FIXed | SWEep | LIST | COMBined
CW|FIXed
Sets the fixed frequency mode. CW and FIXed are synonyms.
The instrument operates at a defined frequency, set with com-
mand [:SOURce<hw>]:FREQuency[:CW|FIXed].
SWEep
Sets sweep mode.
The instrument processes frequency (and level) settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:FREQuency:STARt on page 1132 and [:
SOURce<hw>]:FREQuency:STOP on page 1132,
[:SOURce<hw>]:FREQuency:CENTer on page 1131,
[:SOURce<hw>]:FREQuency:SPAN on page 1131,
[:SOURce<hw>]:FREQuency:MANual on page 1130
LIST
Sets list mode.
The instrument processes frequency and level settings by
means of values loaded from a list.
To configure list mode settings, use the commands of the Chap-
ter 14.19.12, "SOURce:LIST Subsystem", on page 1164.
COMBined
Sets the combined RF frequency / level sweep mode.
The instrument processes frequency and level settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:COMBined:FREQuency:STARt
on page 1213 and [:SOURce<hw>]:COMBined:FREQuency:
STOP on page 1214,
[:SOURce<hw>]:COMBined:POWer:STARt on page 1214
and [:SOURce<hw>]:COMBined:POWer:STOP on page 1214
*RST:
CW
Example: See Example "Frequency configuration" on page 1124,
and Example "Setup an RF frequency or power sweep"
on page 1206
SOURce Subsystem