Remote Control Commands
R&S
®
SMW200A
1129User Manual 1175.6632.02 ─ 30
Manual operation: See "State (RF frequency sweep)" on page 509
See "State (RF frequency / level sweep)" on page 509
See "State" on page 518
[:SOURce<hw>]:FREQuency[:CW|FIXed] <Fixed>
Sets the frequency of the RF output signal in the selected path.
The effect depends on the selected mode:
In CW mode (FREQ:MODE CW | FIXed), the instrument operates at a fixed fre-
quency.
In sweep mode (FREQ:MODE SWE), the value applies to the sweep frequency.
The instrument processes the frequency settings in defined sweep steps.
In user mode (FREQ:STEP:MODE USER), you can vary the current frequency step
by step.
Parameters:
<Fixed> float
The following settings influence the value range:
An offset set with the command [:SOURce<hw>]:FREQuency:
OFFSet
Numerical value
Sets the frequency in CW and sweep mode
UP|DOWN
Varies the frequency step by step in user mode.
The frequency is increased or decreased by the value set with
the command [:SOURce<hw>]:FREQuency:STEP[:
INCRement].
Range:
(RFmin + OFFSet) to (RFmax + OFFSet)
*RST: 100 MHz
Example: See Example "Frequency configuration" on page 1124
Example: See Example "Setup an RF frequency or power sweep"
on page 1206
Options: RFmax depends on the installed options, for example 3 GHz
with R&S SMW-B1003
Manual operation: See "Frequency" on page 462
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL <Rcl>
Set whether the RF frequency value is retained or taken from a loaded instrument con-
figuration, when you recall instrument settings with command *RCL.
The selected mode applies to both RF paths, i.e. a specified suffix is ignored.
Parameters:
<Rcl> INCLude | EXCLude
SOURce Subsystem