EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Physical Location of the Input and Output Interfaces

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Signal Routing and System Configuration
R&S
®
SMW200A
114User Manual 1175.6632.02 ─ 30
put signals and interfaces. Depending on the selected fading and baseband configura-
tion, the digital interfaces can be used as input or output connectors. The different logi-
cal settings of the same physical interface are grouped in different dialogs.
Logical signals, streams and channels at the interfaces
This section provides background information and introduction to the topics. The
description is divided in parts that follow the logical decision process when configuring
the instrument. It supports the manual configuration for simple test scenarios.
The settings provided in the "System Configuration" dialog guide you throughout this
decision process automatically. For complex routing scenarios, like MIMO configuration
and especially configuration involving several instruments, we recommend that you
use the "System Configuration" functionality as a start and central configuration point.
For an overview of the interfaces and their characteristics, refer to:
Chapter 4.3.2, "Overview of the Baseband Signal Sources", on page 116
Chapter 4.3.3, "Overview of the Baseband and RF Output Signals", on page 118
Chapter 4.3.5, "Important Signal Parameters and Interface's Characteristics",
on page 119
4.3.1 Physical Location of the Input and Output Interfaces
This section focuses on the input and output interfaces of the instrument. For a
detailed overview of all front and rear panel connectors, see Chapter 3.2, "Instrument
Tour", on page 37.
Naming conventions for the input and output interfaces and Show Connector
function
The R&S SMW is a highly modular instrument that can be equipped with different com-
binations of hardware and software options. To increase the serviceability of the instru-
ment, the design of the different hardware board is similar. However, this similarity
together with the customizable connectors direction and user definable signals can
cause cabling and configuration faults. To simplify the cabling and to minimize this
error source, it is important to identify the interfaces in any possible hardware and sys-
tem configuration.
The instrument provides the following two methods for identifying the connectors:
A connector naming rule that follows the predefined structure
<Board Name> <Number>[/<Hs Channel>][ <Direction>]
For example, CODER 2 IN or BBMM 2 OUT.
The <Number>/<Hs Channel> designation applies for the HS DIG I/Q connec-
tors, for example BBMM 2/3.
A built-in Show Connector function, that indicates the selected connector on the
rear panel of the instrument with a blinking LED.
The short and long connector names (e.g. COD 2 IN and CODER 2 IN or "Coder 2 IN")
are used interchangeably.
Overview of the Input and Output Signals and Interfaces