Remote Control Commands
R&S
®
SMW200A
1161User Manual 1175.6632.02 ─ 30
Example: See Example "Configuring the LF generator" on page 1153.
Manual operation: See "Period" on page 554
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRIangle:RISE <Rise>
Selects the rise time for the triangle single of the LF generator.
Parameters:
<Rise> float
Range: 1E-6 to 100
Increment: 10E-9
*RST: 0.5E-3
Example: See Example "Configuring the LF generator" on page 1153.
Manual operation: See "Triangle Rise" on page 555
14.19.11.2
LF Sweep Settings
With the commands described in this section, you can configure the sweep of the LF
signal.
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl..................................................1161
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute.............................................. 1161
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:MODE..................................................1162
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:POINts................................................. 1162
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:RETRace..............................................1163
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:RUNNing?............................................ 1163
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SHAPe.................................................1163
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SPACing...............................................1163
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic................................ 1164
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear].......................................1164
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl <Dwell>
Sets the dwell time for each frequency step of the sweep.
Parameters:
<Dwell> float
Range: 0.001 to 100
Increment: 100E-6
*RST: 0.01
Default unit: s
Example: See Example "Setup an LF sweep" on page 1152.
Manual operation: See "Dwell Time" on page 512
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute
Immediately starts an LF sweep.