Remote Control Commands
R&S
®
SMW200A
1179User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:LIST:DEXChange:MODE <Mode>
Determines the import or export of a list.
Specify the source or destination file with the command [:SOURce<hw>]:LIST:
DEXChange:SELect.
Parameters:
<Mode> IMPort | EXPort
*RST: IMPort
Example: See Example "List mode data exchange" on page 1167.
Manual operation: See "Mode" on page 522
[:SOURce<hw>]:LIST:DEXChange:SELect <Filename>
Selects the ASCII file for import or export, containing a list.
Parameters:
<Filename> string
Filename or complete file path; file extension can be omitted.
Example: See Example "List mode data exchange" on page 1167.
Manual operation: See "Select Source/Select ASCII Destination" on page 523
14.19.13 SOURce:NOISe Subsystem
The SOURce:NOISe subsystem contains the commands for:
Setting the noise modulation signal. The noise generator is optional.
Generating impulsive and phase noise.
R&S SMW-K810
Example: Configuring the noise generator
SOURce1:NOISe:DISTribution GAUSs
SOURce1:NOISe:BANDwidth 10000000
SOURce1:NOISe:BANDwidth:STATe 1
SOURce1:LFOutput1:SOURce NOIS
SOURce1:LFOutput1:STATe 1
SOURce1:NOISe:LEVel1:RELative?
SOURce1:NOISe:LEVel1:ABSolute?
SOURce Subsystem