Remote Control Commands
R&S
®
SMW200A
1183User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:NOISe:IMPulsive:FRAMe <IplsFrameDur>
Sets the time intervals at which the bursts occur.
Parameters:
<IplsFrameDur> float
Range: 0.1E-3 to 1
Increment: 0.1E-3
*RST: 0.01
Example: See Example "Generating impulsive noise" on page 1180.
Manual operation: See "Frame Duration" on page 436
[:SOURce<hw>]:NOISe:IMPulsive:PULSe <IplsPulse>
Sets the number of noise pulses per burst.
Parameters:
<IplsPulse> integer
Range: 1 to 65535
*RST: 1
Example: See Example "Generating impulsive noise" on page 1180.
Manual operation: See "Pulses per Burst" on page 436
[:SOURce<hw>]:NOISe:IMPulsive:MINSpace <IplsMinSpace>
[:SOURce<hw>]:NOISe:IMPulsive:MAXSpace <IplsSpaceMax>
If more than 1 pulse per burst are enabled ([:SOURce<hw>]:NOISe:IMPulsive:
PULSee.g. 2), sets the minimum/maximum spacing between 2 pulses within a burst.
Parameters:
<IplsSpaceMax> float
Range: 0.25E-6 to 0.01
Increment: 0.01E-6
*RST: 0.25E-6
Example: See Example "Generating impulsive noise" on page 1180.
Manual operation: See "Pulse Spacing Maximum" on page 437
[:SOURce<hw>]:NOISe:IMPulsive[:BURSt]:DURation?
Queries the time during which the noise generator is active in a frame.
Return values:
<IplsBurstDur> float
Range: 0.01E-6 to 1
Increment: 0.01E-6
*RST: n.a. (no preset. default: 0.01)