Remote Control Commands
R&S
®
SMW200A
1185User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:NOISe:PHASenoise:SHAPe:STORe <PhaseNoiseStore>
Saves the current SSB profile settings into the selected file; the file extension (*.fcf)
is assigned automatically.
Parameters:
<PhaseNoiseStore> "<filename>"
Filename or complete file path
Example: See Example "Generating phase noise" on page 1180.
Manual operation: See "Load Profile, Save Profile" on page 439
[:SOURce<hw>]:NOISe:PHASenoise:LEVel<ch> <PhaseNoiseLevel>
Sets the level value of the points, where the points are designated by the suffix <ch>.
Suffix:
<ch>
.
1 to 5
Defines the frequency/noise point number
Parameters:
<PhaseNoiseLevel> float
Range: -199.99 to 0
Increment: 0.01
*RST: 0
Example: See Example "Generating phase noise" on page 1180.
Manual operation: See "Phase Noise Table" on page 439
[:SOURce<hw>]:NOISe:PHASenoise:FREQuency<ch> <PhaseNoiseFreq>
Sets the frequency value of the points, where the points are designated by the suffix
<ch>.
Suffix:
<ch>
.
1 to 5
Defines the frequency/noise point number
Parameters:
<PhaseNoiseFreq> integer
Range: 10 to 10E6
*RST: 0
Example: See Example "Generating phase noise" on page 1180.
Manual operation: See "Phase Noise Table" on page 439
SOURce Subsystem