Remote Control Commands
R&S
®
SMW200A
1189User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:POWer:ALC:MODE?
Queries the currently set ALC mode.
See [:SOURce<hw>]:POWer:ALC[:STATe] on page 1189.
Return values:
<PowAlcMode> 0 | AUTO | 1 | PRESet | OFFTable | ON | OFF | ONSample |
ONTable
Example:
POW:ALC:MODE?
Response: ONTable "Table and On" is set automatically.
Usage: Query only
Manual operation: See "Mode" on page 568
[:SOURce<hw>]:POWer:ALC[:STATe]
<State>
Activates automatic level control in the selected mode.
How to: See Chapter 8.9.2.3, "How to Enable the ALC", on page 570.
Parameters:
<State> AUTO | OFFTable | ON | ONSample | ONTable | OFF
AUTO
Adjusts the output level to the operating conditions automati-
cally.
OFFTable
Controls the level with the attenuation values of the internal ALC
table.
ON
Activates internal level control permanently.
OFF
Deactivates internal level control, "Sample & Hold" mode is
active.
ONSample
Starts the internal level control with the first change.
ONTable
Starts with the attenuation setting from the table and continues
with automatic level control.
*RST:
AUTO
For more details on the individual settings, an overview of the
functionality and details on what is to be considered, see "ALC
states and their effects" on page 566.
Example:
POW:ALC ON
Activates internal level control.
Manual operation: See "State" on page 568
SOURce Subsystem