Remote Control Commands
R&S
®
SMW200A
1193User Manual 1175.6632.02 ─ 30
Example:
SOURce1:POWer:LBEHaviour AUTO
Manual operation: See "Setting Characteristics" on page 468
[:SOURce<hw>]:POWer:LIMit[:AMPLitude] <Amplitude>
Limits the maximum RF output level in CW and sweep mode.
It does not influence the "Level" display or the response to the query [:
SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude].
Parameters:
<Amplitude> float
Range: depends on the installed options
Increment: 0.01
*RST: n.a. (factory preset: 30)
Example:
SOURce1:POWer:LIMit:AMPLitude 10
Manual operation: See "Limit" on page 470
[:SOURce<hw>]:POWer:LMODe <LevMode>
Sets the RF level mode.
Parameters:
<LevMode> NORMal | LOWNoise | LOWDistortion
NORMal
Supplies the RF signal with the standard power level of the
instrument.
LOWNoise
Supplies a very low noise sinewave signal.
LOWDistortion
Supplies a very pure sinewave signal.
*RST: NORMal
Example:
SOURce1:POWer:LMODe LOWD
Sets low distortion mode. The instrument reduces distortions of
the RF signal to a minimum.
Manual operation: See "Mode" on page 468
[:SOURce<hw>]:POWer:MANual <Manual>
Sets the level for the subsequent sweep step if SWE:POW:MODE MAN.
Use a separate command for each sweep step.
Parameters:
<Manual> float
You can select any level within the setting range, where:
STARt is set with [:SOURce<hw>]:POWer:STARt
SOURce Subsystem