Remote Control Commands
R&S
®
SMW200A
1194User Manual 1175.6632.02 ─ 30
STOP is set with [:SOURce<hw>]:POWer:STOP
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:
IMMediate]:OFFSet
Range: (STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01
Default unit: dBm
Example: See Example "Setup an RF frequency or power sweep"
on page 1206
Manual operation: See "Current Level" on page 510
[:SOURce<hw>]:POWer:MODE <Mode>
Selects the operating mode of the instrument to set the output level.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Operates at a constant level.
CW and FIXed are synonyms.
To set the output level value, use the command [:
SOURce<hw>]:POWer[:LEVel][:IMMediate][:
AMPLitude].
SWEep
Sets sweep mode.
Set the range and current level with the commands:
[:SOURce<hw>]:POWer:STARt and [:SOURce<hw>]:
POWer:STOP,
[:SOURce<hw>]:POWer:MANual.
*RST:
CW
Example: See Example "Setup an RF frequency or power sweep"
on page 1206
Manual operation: See "State (RF level sweep)" on page 509
See "State (RF frequency / level sweep)" on page 509
[:SOURce<hw>]:POWer:PEP?
Queries the PEP "Peak Envelope Power) of digital modulation or digital standards at
the RF output. This value corresponds to the level specification, displayed in the status
bar (header).
Return values:
<Pep> float
Example:
POW:PEP?
Response: "4"
The PEP value of digital modulation is 4 dBm at the RF output.
SOURce Subsystem