Remote Control Commands
R&S
®
SMW200A
1195User Manual 1175.6632.02 ─ 30
Usage: Query only
[:SOURce<hw>]:POWer:POWer <Power>
Sets the level at the RF output connector.
This value does not consider a specified offset.
The command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
sets the level of the "Level" display, that means the level containing offset.
See Chapter 8.4.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 472.
Parameters:
<Power> float
Level at the RF output, without level offset
Range: See data sheet
Increment: 0.01
Default unit: dBm
Example:
SOURce1:POWer:POWer 15
Sets the level at RF output
Manual operation: See "Amplitude" on page 467
[:SOURce<hw>]:POWer:STARt <Start>
[:SOURce<hw>]:POWer:STOP <Stop>
Sets the RF start/stop level in sweep mode.
Parameters:
<Stop> float
Sets the setting range calculated as follows:
(Level_min + OFFSet) to (Level_max + OFFSet)
Where the values are set with the commands:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
[:SOURce<hw>]:POWer:STARt
[:SOURce<hw>]:POWer:STOP
Range: Minimum level to maximum level
*RST: -30 (Start)/ -10 (Stop)
Default unit: dBm
Example: See Example "Setup an RF frequency or power sweep"
on page 1206
Manual operation: See "Start Level / Stop Level" on page 515
[:SOURce<hw>]:POWer:STEP:MODE <Mode>
Defines the type of step width to vary the RF output power step-by-step with the com-
mands POW UP or POW DOWN.
SOURce Subsystem