Remote Control Commands
R&S
®
SMW200A
1197User Manual 1175.6632.02 ─ 30
Parameters:
<Offset> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Default unit: dB
Level offset is always expreced in dB; linear units (V, W, etc.) are
not supported
Example:
POWer:OFFSet 10
Sets the RF level offset to 10 dB
Manual operation: See "Offset" on page 470
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL <Rcl>
Determines whether the current level is retained or if the stored level setting is adopted
when an instrument configuration is loaded.
The setting is valid for both paths. If a suffix is specified, it is ignored.
Parameters:
<Rcl> INCLude | EXCLude
INCLude
Takes the current level when an instrument configuration is loa-
ded.
EXCLude
Retains the current level when an instrument configuration is
loaded.
*RST: INCLude
Example:
POW:RCL INCL
Takes the level value from an instrument configuration loaded
with command *RCL.
Manual operation: See "Exclude Level" on page 683
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude] <Amplitude>
Sets the RF level applied to the DUT.
To activate the RF output use command :OUTPut<hw>[:STATe] ("RF On"/"RF Off").
The following applies POWer = RF output level + OFFSet, where:
●
POWer is the values set with [:SOURce<hw>]:POWer[:LEVel][:
IMMediate][:AMPLitude]
●
RF output level is set with [:SOURce<hw>]:POWer:POWer
●
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
Parameters:
<Amplitude> float
SOURce Subsystem