Remote Control Commands
R&S
®
SMW200A
1198User Manual 1175.6632.02 ─ 30
The following settings influence the value range:
OFFSet set with the command [:SOURce<hw>]:POWer[:
LEVel][:IMMediate]:OFFSet
Numerical value
Sets the level
UP|DOWN
Varies the level step by step.
The level is increased or decreased by the value set with the
command [:SOURce<hw>]:POWer:STEP[:INCRement].
Range: (Level_min + OFFSet) to (Level_max + OFFStet)
*RST: -30
Default unit: dBm
Example:
POWer -30
Sets the RF level
Example: See also [:SOURce<hw>]:POWer:STEP:MODE on page 1195.
Manual operation: See "Amplitude" on page 467
[:SOURce<hw>]:POWer:RANGe:LOWer?
[:SOURce<hw>]:POWer:RANGe:UPPer?
Queries the current interruption-free range of the level.
Return values:
<Upper> float
Increment: 0.01
Default unit: dBm
Example:
SOURce1:POWer:RANGe:UPPer?
// -15
SOURce1:POWer:RANGe:LOW?
// -50
Usage: Query only
Manual operation: See "Level Range" on page 517
14.19.17 SOURce:ROSCillator Subsystem
The SOURce:ROSCillator subsystem contains the commands for setting the exter-
nal and internal reference frequency.
The commands of this subsystem are not affected by an instrument reset (*RST
on page 847).
SOURce Subsystem