Remote Control Commands
R&S
®
SMW200A
1202User Manual 1175.6632.02 ─ 30
[:SOURce]:ROSCillator:EXTernal:NSBandwidth?
Queries the nominal synchronization bandwidth for the selected external reference fre-
quency and synchronization bandwidth.
Return values:
<NomBandwidth> string
Example:
SOUR:ROSC:EXT:NSB?
Usage: Query only
Manual operation: See "Nominal Synchronization Bandwidth" on page 476
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE
<OutpFreqMode>
Selects the mode for the determination and output of the reference frequency.
Parameters:
<OutpFreqMode> DER10M | SAME
*RST: n.a. (factory preset: DER10M)
Example: See Example "Configuring the reference oscillator"
on page 1199.
Manual operation: See "Reference Output/1 GHz Reference Output" on page 477
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue <Value>
Specifies the frequency correction value (adjustment value).
Parameters:
<Value> integer
*RST: 0
Example: See [:SOURce]:ROSCillator[:INTernal]:ADJust[:
STATe] on page 1202
Manual operation: See "Adjustment Value" on page 479
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe] <State>
Determines whether the calibrated (off) or a user-defined (on) adjustment value is
used for fine adjustment of the frequency.
Parameters:
<State> 0 | 1 | OFF | ON
0
Fine adjustment with the calibrated frequency value
SOURce Subsystem