Remote Control Commands
R&S
®
SMW200A
1208User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute............................................................ 1212
[:SOURce<hw>]:SWEep:POWer:RETRace.....................................................................1212
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace............................................................1212
[:SOURce<hw>]:SWEep:POWer:RUNNing?................................................................... 1213
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?.......................................................... 1213
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic.............................................. 1213
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear].....................................................1213
[:SOURce<hw>]:COMBined:FREQuency:STARt............................................................. 1213
[:SOURce<hw>]:COMBined:FREQuency:STOP..............................................................1214
[:SOURce<hw>]:COMBined:POWer:STARt.................................................................... 1214
[:SOURce<hw>]:COMBined:POWer:STOP.....................................................................1214
[:SOURce<hw>]:SWEep:COMBined:COUNt...................................................................1215
[:SOURce<hw>]:SWEep:COMBined:DWELl................................................................... 1215
[:SOURce<hw>]:SWEep:COMBined:EXECute................................................................1215
[:SOURce<hw>]:SWEep:COMBined:MODE....................................................................1216
[:SOURce<hw>]:SWEep:COMBined:RETRace............................................................... 1216
[:SOURce<hw>]:SWEep:COMBined:SHAPe...................................................................1216
[:SOURce<hw>]:SWEep:RESet[:ALL]............................................................................ 1217
[:SOURce<hw>]:SWEep:POWer:AMODe <AMode>
Selects the power attenuator mode for the level sweep.
Parameters:
<AMode> NORMal | HPOWer
NORMal
Performs the level settings in the range of the built-in attenuator.
HPOWer
Performs the level settings in the high level range.
*RST: NORMal(HighPower)|AUTO
Example:
SWE:POW:AMOD HPOW
Selects the high level ranges for level sweep.
[:SOURce<hw>]:SWEep:POWer:DWELl <Dwell>
Sets the dwell time for a level sweep step.
Parameters:
<Dwell> float
Range: 0.001 to 100
Increment: 100E-6
*RST: 0.01
Example: See Example "Setup an RF frequency or power sweep"
on page 1206.
Manual operation: See "Dwell Time" on page 512
SOURce Subsystem