Remote Control Commands
R&S
®
SMW200A
1212User Manual 1175.6632.02 ─ 30
Example: See Example "Setup an RF frequency or power sweep"
on page 1206.
Manual operation: See "Spacing" on page 511
[:SOURce<hw>]:SWEep:POWer:SHAPe <Shape>
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe <Shape>
Determines the waveform shape for a frequency sweep sequence.
Parameters:
<Shape> SAWTooth | TRIangle
*RST: SAWTooth
Example: See Example "Setup an RF frequency or power sweep"
on page 1206.
Manual operation: See "Shape" on page 511
[:SOURce<hw>]:SWEep:POWer:EXECute
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
Executes an RF frequency sweep.
The command performs a single sweep and is therefore only effective in manual
sweep mode.
Example:
See Example "Setup an RF frequency or power sweep"
on page 1206.
Usage: Event
Manual operation: See "Execute Single Sweep" on page 512
[:SOURce<hw>]:SWEep:POWer:RETRace <State>
[:SOURce<hw>]:SWEep[:FREQuency]:RETRace <State>
Activates that the signal changes to the start frequency value while it is waiting for the
next trigger event.
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single".
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Setup an RF frequency or power sweep"
on page 1206.
Manual operation: See "Retrace" on page 511