Remote Control Commands
R&S
®
SMW200A
1228User Manual 1175.6632.02 ─ 30
Parameters:
<Address> integer
Range: 0 to 30
*RST: 28
Example: See Example "Finding out the used VISA ressource strings"
on page 1217.
:SYSTem:COMMunicate:HISLip:RESource?
Queries the VISA resource string. This string is used for remote control of the instru-
ment with HiSLIP protocol.
Return values:
<Resource> string
Example: See Example "Finding out the used VISA ressource strings"
on page 1217.
Usage: Query only
:SYSTem:COMMunicate:NETWork:IPADdress <IpAddress>
Sets the IP address.
Parameters:
<IpAddress> string
Range: 0.0.0.0. to ff.ff.ff.ff
Example: See Example "Retrieving information on network-related set-
tings" on page 1217.
Manual operation: See "IP Address" on page 787
:SYSTem:COMMunicate:NETWork:IPADdress:MODE <Mode>
Selects manual or automatic setting of the IP address.
Parameters:
<Mode> AUTO | STATic
*RST: n.a. (factory preset: AUTO)
Example: See Example "Retrieving information on network-related set-
tings" on page 1217.
Example:
SYSTem:COMMunicate:NETWork:IPADdress:MODE
STATic
SYSTem:COMMunicate:NETWork:IPADdress
"10.113.0.105"
Manual operation: See "Address Mode" on page 787