17.3.1.2 How To Set Data and Time
To select the timezone
1. Press the [SETUP] key.
2. Select "Maintenance > Date /Time > Timezone".
3. Select continent and city of your location.
Tip: By typing the first letter, you can quickly navigate through the lists to find the
desired destination.
4. Close the dialogs.
The instrument adjusts the time according to the selected location.
To set the date and time
1. Press the [SETUP] key.
2. Select "Security > Protection".
3. Enable the "Protection Level 1".
The default password is 123456.
4. Select "Setup > Maintenance > Date / Time".
5. Adjust the settings.
6. Close the dialogs.
The instrument adopts the new date and time.