Maintenance, Storage and Disposal
R&S
®
SMW200A
1282User Manual 1175.6632.02 ─ 30
Continue Adjustment on Error
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
This function is password-protected. Unlock the protection level 1 to access it.
Remote command:
:CALibration<hw>:CONTinueonerror on page 866
Log Debug Info
Activates logging of the internal adjustments.
This function is password-protected. Unlock the protection level 1 to access it.
Remote command:
:CALibration<hw>:DEBug on page 866
17.3.3.2 How to Use the Internal Adjustments
Deciding whether to run internal adjustments
1. Select "System Config > Setup > Internal Adjustment".
2. Observe the status and color indication in the section "Since Last Full Adjustment".
Green: Internal adjustments are not required.
Red: Internal adjustments are required. Observe also the indication in the "Informa-
tion" field.
3. We recommend that you run internal adjustments in the following cases:
Before starting any application, that requires a maximum of level accuracy
When a long period of time has passed since the last adjustments
If the ambient temperature of the instrument significantly differs from the one of
the last adjustments
4. Proceed as described in "Running internal adjustments" on page 1282.
Running internal adjustments
1.
NOTICE! Adjustments can be invalid if performed when the instrument is not
warmed-up.
Wait until the instrument has reached its operating temperature before you start the
adjustment procedure.
The warm-up time is up to 30 minutes.
2. Select "System Config > Setup > Internal Adjustment > Adjust All".
In two-path instruments, you can adjust each path separately.
The adjustment process starts.
3. Do not interrupt the adjustment process.
The extent of the adjustments depends on the installed options.
A progress indicator shows the status of the adjustment process.
Performing Maintenance Tasks