Maintenance, Storage and Disposal
R&S
®
SMW200A
1283User Manual 1175.6632.02 ─ 30
If any error occurs, the process aborts and an error message appears in the "Info"
line.
If there is an error:
Select "System Config > Info".
Observe the error message.
Continue the process.
See "Continuing the adjustment process if error appears" on page 1283.
Record the results in a log file.
Select "System Config > Setup > Internal Adjustment > Adjustment Configura-
tion" > Log Debug Info > "On".
Send this report to the Rohde & Schwarz customer support for an initial check.
See Chapter 15.8, "Collecting Information for Technical Support",
on page 1267.
Continuing the adjustment process if error appears
Per default, if any error occurs during the adjustment process, the process aborts. An
error message appears in the "Info" line.
If you prefer to finish the adjustments also if there is an error, proceed as follows:
1. Select "System Config > Setup > Security > Protection" an unlock the protection
level 1.
See "Protection Level/Password" on page 754.
2. Select "System Config > Setup > Internal Adjustment > Configuration > Continue
Adjustment on Error > On".
3. Proceed as described in "Running internal adjustments" on page 1282.
17.3.4 Selftest
Self-test is provided for service purposes.
Performing Maintenance Tasks

Table of Contents