Signal Routing and System Configuration
R&S
®
SMW200A
165User Manual 1175.6632.02 ─ 30
Remote command:
[:SOURce<hw>]:BBIN:POWer:PEAK on page 963
Level
Indicates the estimated RMS level acquired with "Auto Level Set".
Remote command:
[:SOURce<hw>]:BBIN:POWer:RMS? on page 963
4.6.1.4 Signal Monitoring Settings
Option: R&S SMW-B10
Access:
Select "BB Input > Digital I/Q In > Baseband Input Settings > Signal Monitor".
Settings:
Signal Monitoring........................................................................................................ 165
Signal Monitoring
The R&S SMW monitors constantly the input signal (crest factor and peak level) and
indicates an overflow status, if the selected "Peak Level" (in dB full scale) value does
not correspond with the real value.
Note: The overflow status is indicated as long as the "Peak Level" value is not correc-
ted.
To avoid overflow effects, enable automatic measurement with the "Auto Level Set"
function.
"Overflow/No Overflow"
Indicates whether the input is overloaded or not. An additional indica-
tion is displayed in the block diagram close to the corresponding con-
nector.
Remote command:
[:SOURce<hw>]:BBIN:OLOad:STATe? on page 963