Signal Routing and System Configuration
R&S
®
SMW200A
171User Manual 1175.6632.02 ─ 30
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:SRATe on page 959
Crest Factor ← Channel Table
If DIG IQ Auto Setting > "Off", sets the crest factor per channel.
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:CFACtor on page 958
Peak Level ← Channel Table
If DIG IQ Auto Setting > "Off", sets the peak level per channel. The value is set as
attenuation in digital baseband domain.
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:PEAK on page 958
Level ← Channel Table
Indicates the level, set automatically depending on the selected peak level and crest
factor.
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:POWer:RMS on page 959
BB ← Channel Table
Activates the channel. Only one channel can be activated at a time. Toggle the state of
the channel to test all of them subsequently.
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:BB:STATe on page 957
Aggregated Link Sample Rate
Indicates the aggregated sample rate and the maximum sample rate, where:
●
The aggregated sample rate is the sum of the sample rates of all active channels.
It cannot exceed the maximum sample rate.
●
The R&S SMW supports two standard max. sample rate values.
This parameter indicates the currently selected one, where the max. sample rate is
selected depending on the capabilities of the transmitter/receiver.
For more information, see data sheet.
Remote command:
[:SOURce<hw>]:BBIN:SRATe:SUM? on page 960
[:SOURce<hw>]:BBIN:SRATe:MAX? on page 961
DIG IQ Auto Setting
Activates an automatic adjustment of the baseband input signal.
The R&S SMW receives peak level, level and crest factor values directly from the con-
nected transmitter and recognizes changes automatically.
Remote command:
[:SOURce<hw>]:BBIN:DIGital:ASETting:STATe on page 961
Measurement Period
For CODER connectors:
Digital Baseband Input Settings