Signal Routing and System Configuration
R&S
®
SMW200A
190User Manual 1175.6632.02 ─ 30
Observe the preview diagram in the "Fading/Baseband Settings" dialog.
This diagram provides a short description and an example of the common applica-
tion or test scenario of the selected configuration.
For description of the further steps, see:
Chapter 4.9.3, "How to Define the MIMO Scenario", on page 191
Chapter 4.9.4, "How to Route the I/Q Streams to the Output Interfaces",
on page 193
See also:
Chapter 4.9.5, "How to Cable the Instruments in MIMO Test Setups",
on page 194
Chapter 4.9.6, "How to Connect and Configure External Instruments",
on page 195
4.9.2 How to Find Out where the Input/Output Interface Is Located
See:
"To find out the correct connector" on page 190
"To find out the connector's direction" on page 191
To find out the correct connector
1. Select "Taskbar > System Configuration > External RF and I/Q".
2. In the overview table, navigate to the connector an external instrument has to be
connected to (e.g. "BBMM 1").
Select "External Instruments > Config".
3. In the "External Instrument Configuration" dialog, select "Show Connector".
A blinking LED on the rear of the instrument indicates the connector's position.
How to Connect External Instruments and Configure the Signal Flow