EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - How to Cable the Instruments in MIMO Test Setups

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Signal Routing and System Configuration
R&S
®
SMW200A
194User Manual 1175.6632.02 ─ 30
Tip: To route more streams to one output connector, in the"I/Q Stream Mapper"
dialog and for the particular connector, select "Combination> Add" and select the
streams to be routed to it.
4.9.5 How to Cable the Instruments in MIMO Test Setups
The MxN MIMO test setups with more than two Rx antennas (N>2) require the signal
of at least three antennas. The R&S SMW generates the required signal and outputs
two of the antennas signals at the RF outputs. The others signals can be routed to the
provided analog and digital I/Q connectors and then processed by a connected instru-
ment, like an RF frontend. Several RF frontend instruments can serve as extended
RFs for the R&S SMW, for example R&S SGS, or R&S SGT.
The Figure 4-22 shows a 1x4x4 MIMO test setup with two R&S SGS connected to the
analog I/Q outputs of the R&S SMW.
For information on how to connect instruments to the DIG I/Q interfaces, see Fig-
ure 4-24.
How to Connect External Instruments and Configure the Signal Flow

Table of Contents