Signal Routing and System Configuration
R&S
®
SMW200A
199User Manual 1175.6632.02 ─ 30
Figure 4-24: Example of test setup: 1x4x4 MIMO with R&S
SMW and 2xR&S
SGT connected to
the BBMM1/2.
Blue circle = front panel connector
Gray circle = rear panel connector
2. Follow the steps described in "To connect instruments to the I/Q analog interface"
on page 196.
a) In the "System Config > System Configuration > External RF und I/Q", enable
"Auto Connect > On".
b) Use the example initialization file shown in "To connect instruments to the I/Q
analog interface" on page 196.
The "System Configuration > External RF and I/Q" dialog confirms a correct and
successful configuration by the active "Rem Conn" status and active "I/Q Conn"
status indication.