Signal Routing and System Configuration
R&S
®
SMW200A
204User Manual 1175.6632.02 ─ 30
To create an SCPI sequence or your own initialization sequence and send it to an R&S
SGS connected to the I/Q interface, proceed as follows:
1. On a remote computer, open the notepad or other similar tool.
Write a sequence of SCPI commands.
2. Store the SCPI sequence in a file with file extension *.iec.
3. Use one of the standard file transfer methods to transfer the file to the user direc-
tory /usb/ of the R&S SMW.
See Chapter 11.9, "How to Transfer Files from and to the Instrument",
on page 696.
4. Select "System Config > System Configuration > External RF und I/Q".
5. Select "I/Q OUT 1 > External Instrument > Config"
6. Select "External Instrument Configuration > Remote Control":
a) Select "Mode > Sequence".
b) Select "File" and select the *.iec file form the user directory /usb/.
c) Select "Send Sequence".
7. Alternativel:
a) Select "External Instrument Configuration > Initialization Sequence".
b) Select "Initialization Sequence" and select the *.iec file form the user direc-
tory /usb/.
See also:
●
Example "Controlling the R&S SMW and external RF instruments with the same
application program" on page 908
●
Example "Connecting an external instrument and sending SCPI commands to it"
on page 911
4.9.7 How to Generate a 2x8 MIMO Signal for BS Tests
Refer to Table 4-1 to find out the suitable system configuration and the required addi-
tional instruments for this particular test.
There are two suitable test setups:
●
A test setup with one R&S SMW and six R&S SGT
See "Test setup with R&S SMW and 6xR&S SGT" on page 204
●
A test setup with multiplexed digital output signals
See user manual R&S SMW-K551 Generation of Digital "Slow IQ" Signals.
Test setup with R&S SMW and 6xR&S SGT
This test setup is a logical extension of the 1x4x4 test case.
How to Connect External Instruments and Configure the Signal Flow