EasyManua.ls Logo

Rohde & Schwarz R&S SMW200A - Page 213

Rohde & Schwarz R&S SMW200A
1403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Signal Routing and System Configuration
R&S
®
SMW200A
213User Manual 1175.6632.02 ─ 30
8. Select "Baseband Input Settings > General > State > On" to enable the processing
of the external digital signal.
9. Open "Baseband Input Settings > Signal Monitoring" and monitor the signal over-
flow status.
To apply an external analog signal directly to the I/Q modulator
Fed the external analog wideband signal at the I/Q input connectors (see Table 4-3).
The signal is used as an input signal for the I/Q modulator.
Figure 4-32: Example of test setup (simplified)
1. To configure the input signal source of the I/Q modulator, select "I/Q Mod > I/Q Mod
In > Analog Wideband I/Q In".
2. To access the dialog box for configuring the I/Q modulator settings, select "I/Q Mod
> I/Q Settings > I/Q Modulator".
Refer to Chapter 7, "Applying I/Q Vector Modulation", on page 448 for a description
of the settings.
3. In the "I/Q Modulator" dialog, select "General > Analog Wideband I/Q Input > Crest
Factor".
Adjust the crest factor and gain values.
4. To impair the analog signal, adjust the "I/Q Modulator > Impairments" settings (see
also Chapter 6.4, "Impairing the Signal", on page 439).
5. Select "I/Q Mod > On" or "I/Q Mod > I/Q Modulator > General > State > On" to
enable the processing of the external analog signal.
To apply an external differential analog signal directly to the I/Q modulator
1. Select "I/Q Mod > I/Q Settings > Source > Differential Analog I/Q Input".
2. Follow the steps described in "To apply an external analog signal directly to the I/Q
modulator" on page 213.
How to Connect External Instruments and Configure the Signal Flow

Table of Contents