Configuring the Internal Baseband Source
R&S
®
SMW200A
407User Manual 1175.6632.02 ─ 30
5. Adjust further settings, e.g. trigger setting, as required.
Enable signal generation with "Multi-Carrier Continuous Wave > State > ON".
6. Enable the signal output with "RF > ON".
7. On the connected signal analyzer, perform the required measurement tasks. Mea-
sure the frequency response at the outputs of the DUT.
5.12.5 References
Table 5-27: MCCW default values
Parameter Value
State Not affected by "Set to Default"
Carrier Setup
Number of Carriers 64
Carrier Spacing 10 kHz
Optimize Crest Factor Chirp
Desired Crest Factor 3 dB
Trigger
Mode Auto
Source Internal
Ext. Delay 0
Ext. Inhibit 0
Marker
Channel 1...4 Restart
Clock
Generating Multi-Carrier Continuous Wave Signal