Configuring the RF Signal
R&S
®
SMW200A
463User Manual 1175.6632.02 ─ 30
The display has been disabled for security reasons.
See:
Annotation Frequency
Annotation Amplitude
The display is disabled when:
List mode is running, see Chapter 8.7, "Varying the RF Signal in List or Sweep
Mode", on page 491.
Digital only signals are generated, see the user manual "R&S SMW-K551 Gen-
eration of Digital "Slow IQ" Signals"
Remote command:
[:SOURce<hw>]:FREQuency[:CW|FIXed] on page 1129
Offset
Sets the frequency offset.
This value represents the frequency shift of a downstream instrument, like for example
a mixer.
The "Frequency" value displayed in the status bar is the resulting frequency, as it is at
the output of the downstream instrument. The frequency at the R&S SMW RF output is
not changed.
See Chapter 8.4.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 472.
Remote command:
[:SOURce<hw>]:FREQuency:OFFSet on page 1131
Multiplier
Sets the multiplication factor for the RF frequency.
This value represents the multiplication factor of a downstream instrument, as for
example a multiplier. You can also assign a negative multiplication factor, for example
-1.0 to support frequency converters working in the reverse frequency position.
The "Frequency" value displayed in the status bar is the resulting frequency, as it is at
the output of the downstream instrument. The frequency at the R&S SMW RF output is
not changed.
See Chapter 8.4.4.1, "Displayed RF Frequency and Level Values with Downstream
Instruments", on page 472.
Remote command:
[:SOURce<hw>]:FREQuency:MULTiplier on page 1130
Converter Mode
Selects the mode for calculating the output frequency of a downstream converter.
Configuring RF Frequency and Level