Configuring the RF Signal
R&S
®
SMW200A
481User Manual 1175.6632.02 ─ 30
Reference Output = "10 MHz, 100 MHz" or "Input Signal (loop through)"
(earlier RF hardware versions: 5 MHz, 10 MHz, 13 MHz)
Set the synchronization bandwidth according to the requirements of the applica-
tion.
Figure 8-3: Synchronizing instruments with a 10 MHz external reference signal
Ref. Frequency Source = e.g., Rohde & Schwarz signal analyzer
f
Ref
= 10 MHz, 100 MHz, 1 MHz to 100 MHz external reference frequency
REF IN, REF OUT = Connectors
You can forward reference frequency between 1 MHz and 100 MHz directly to the out-
put in the same way.
External f
Ref
= 1 MHz to 100 MHz
(1 MHz to 100 MHz at connector REF IN and REF OUT)
Source = "External"
Variable Reference Frequency = "Variable"
External Reference Frequency = current external frequency
Reference Output = "Input Signal (loop through)" or specify the reference fre-
quency the synchronized instrument supports
Set the synchronization bandwidth according to the requirements of the applica-
tion.
Deriving 10 MHz from the external reference frequency
10 MHz reference frequency can be derived from the following external reference sig-
nals:
10 MHz from internal oscillator locked to external input
100 MHz and 1 GHz external reference signals
External reference signal between 1 MHz and 100 MHz
If the external reference signal is interfered (noisy)
Figure 8-4: Synchronizing instruments with 10
MHz (derived from an external reference frequency)
Reference Oscillator