Configuring the RF Signal
R&S
®
SMW200A
510User Manual 1175.6632.02 ─ 30
Note: Active RF frequency, RF level or RF combined sweep modes deactivate the LF
sweep or List mode and vice versa.
Remote command:
[:SOURce<hw>]:LFOutput:FREQuency:MODE on page 1156
Current Frequency
Displays the current frequency sweep value for all sweep Modes, except of Mode >
Step. In this mode, you can enter the next frequency setting manually.
Remote command:
[:SOURce<hw>]:FREQuency:MANual on page 1130
[:SOURce<hw>]:LFOutput:FREQuency:MANual on page 1155
Current Level
Displays the current level sweep value for all sweep Modes, except of Mode > Step. In
this mode, you can enter the next level setting manually.
Remote command:
[:SOURce<hw>]:POWer:MANual on page 1193
Mode
Selects the sweep mode.
See Chapter 8.7.1, "Signal Generation and Triggering in the Sweep and List Modes",
on page 493.
"Auto"
Generates a continuously repeating sweep signal directly after acti-
vating the sweep mode.
The sweep steps are performed automatically, controlled by the dwell
time.
"Single / Extern Single"
Generates a single sweep cycle after a trigger event.
The sweep steps within the cycle are performed automatically, con-
trolled by the dwell time. If one cycle is completed, the instrument
waits for the next trigger event.
"Step / Extern Step"
Generates the sweep signal step-by-step, manually triggered.
"Extern Start/Stop"
Generates a continuously repeating sweep signal that is started,
stopped, and restarted by subsequent external trigger events.
The sweep steps are performed automatically, controlled by the dwell
time.
Remote command:
RF frequency sweep:
[:SOURce<hw>]:SWEep[:FREQuency]:MODE on page 1210
:TRIGger<hw>:FSWeep:SOURce on page 1252
RF level sweep:
[:SOURce<hw>]:SWEep:POWer:MODE on page 1209
:TRIGger<hw>:PSWeep:SOURce on page 1252
RF combined sweep:
[:SOURce<hw>]:SWEep:COMBined:MODE on page 1216
Varying the RF Signal in List or Sweep Mode