Configuring the RF Signal
R&S
®
SMW200A
513User Manual 1175.6632.02 ─ 30
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute on page 1161
:TRIGger<hw>:LFFSweep:IMMediate on page 1254
General:
:TRIGger<hw>[:SWEep][:IMMediate] on page 1254
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
A blinking LED on the front/rear panel indicates the selected connector, too (see Chap-
ter 12.2.2.4, "Status Indicators", on page 718).
Reset Sweep
Resets a sweep.
With the next trigger event, the sweep starts at the initial value.
Remote command:
[:SOURce<hw>]:SWEep:RESet[:ALL] on page 1217
8.7.5.2 Frequency Range Settings
Access:
1. Select for example "RF" > "Sweep/List" > "RF Frequency Sweep"
2. Select for example "Frequency Range".
Table 8-9: Range settings dialog: Examples of RF Frequency and RF Combined Sweep dialogs
> "Frequency Range" > "Frequency / Level Range"
Settings
Start Frequency/Stop Frequency................................................................................ 513
Center Frequency....................................................................................................... 514
Span............................................................................................................................514
Spacing....................................................................................................................... 514
Step Linear/Step Logarithmic......................................................................................514
Start Frequency/Stop Frequency
Defines the frequency sweep range by setting the start and end values.
See Chapter 8.7.2.1, "Correlating Parameters in Sweep Mode", on page 501.
Varying the RF Signal in List or Sweep Mode