Configuring the RF Signal
R&S
®
SMW200A
521User Manual 1175.6632.02 ─ 30
8.7.6.2 List Mode Data Settings
Access:
1. Select "RF" > "Sweep/List" > "List mode".
2. Select "List Mode Data".
This dialog contains the parameters required for creating and editing lists, activat-
ing the learning function and selecting the list processing mode.
Settings
List Mode Data............................................................................................................521
Edit List Mode Data.....................................................................................................521
List Range from/to.......................................................................................................521
List Mode Data
Accesses the standard "Select List" dialog for selecting, creating and editing a list file.
The currently loaded file is indicated.
You can create data lists with the internal editor or import externally created files, see
"Creating and handling lists" on page 504.
Remote command:
[:SOURce<hw>]:LIST:CATalog? on page 1175
[:SOURce<hw>]:LIST:SELect on page 1176
[:SOURce<hw>]:LIST:DELete on page 1175
[:SOURce<hw>]:LIST:DELete:ALL on page 1175
Edit List Mode Data
Opens the editor to insert and save data lists with RF frequency, power and dwell time
values, see Chapter 8.7.7, "List Editor", on page 524.
You find this function also in standard file select dialog, accessed via List Mode Data.
List Range from/to
Defines an index range in the current list by setting the start and stop index.
The instrument generates the signal with the values of the selected index range and
ignores all other list entries.
Remote command:
[:SOURce<hw>]:LIST:INDex:STARt on page 1171
[:SOURce<hw>]:LIST:INDex:STOP on page 1171
Varying the RF Signal in List or Sweep Mode