Configuring the RF Signal
R&S
®
SMW200A
527User Manual 1175.6632.02 ─ 30
To fill the table, select "Fill".
Note: Once you enter a value or fill a column, the editor automatically adds preset val-
ues in the other columns. This functionality protects against data loss, otherwise
incomplete rows will be lost when saving. You can simply override these values.
"From / Range"
Defines the start line and number of lines to be filled.
"Column To Fill"
Selects, if the column is filled up with frequencies in Hz, levels in dBm
or dwell times in s.
"Start Value / End Value"
Sets the start value for frequency, level or dwell time. The end value
is read only and depends on the increment value and the range.
"Increment Value"
Determines the step size.
"Fill"
Fills the column specified in "Column To fill".
8.7.8 How to Generate a Signal in List or Sweep Mode
This section shows you how to configure a varying RF output signal for both, the list
and sweep modes.
To configure the frequency sweep signal
1. In the status bar, enter "Freq = 6 GHz" and "Lev = -25 dBm".
2. Select "RF" > "Sweep/List" > "RF Frequency Sweep"
3. Select "Mode > Auto".
4. Select "Spacing > Linear".
5. Set "Dwell Time > 15.0 ms".
Varying the RF Signal in List or Sweep Mode