Configuring the RF Signal
R&S
®
SMW200A
559User Manual 1175.6632.02 ─ 30
Output Voltage
Sets the peak to peak voltage of the selected LF output source.
This value, e.g. +1 V to -1 V refers to 50 Ohm at the RF output with high termination
impedance at the LF output.
Remote command:
[:SOURce]:LFOutput<ch>:VOLTage on page 1158
DC-Offset
Adds a DC offset to the LF output signal.
Remote command:
[:SOURce]:LFOutput<ch>:OFFSet on page 1157
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
A blinking LED on the front/rear panel indicates the selected connector, too (see Chap-
ter 12.2.2.4, "Status Indicators", on page 718).
Noise Level Settings
Access:
1. Select "RF > Output > LF Output".
2. Set "Source > Noise Generator".
3. Select "Output > Noise Level".
Settings:
Noise Density..............................................................................................................559
Noise Level................................................................................................................. 559
Noise Density
Indicates the level of the noise signal for a bandwidth of 1 Hz (relative).
Remote command:
[:SOURce<hw>]:NOISe:LEVel<ch>:RELative? on page 1181
Noise Level
Indicates the level of the noise signal per Hz within the total bandwidth (absolute).
Analog Modulations